“Medtronic has thoroughly evaluated the Advantage valve to known rigorous standards. The overall clinical performance of the Advantage valve is excellent. Based on reports from clinicians, it is now recognized that the Advantage valve had a rarely expressed characteristic referred to as intermittent regurgitation. Medtronic has received 17 reports of intermittent regurgitation with over 15,000 valves implanted. During the isolated beats when intermittent regurgitation occurs, one of the two leaflets comes very close to closing but does (lees:) not completely close. This characteristic did not manifest itself during pre-clincal testing. Two cases of intermittent regurgitation resulted in misinterpretation of valve performance, leading clinicians, in these two cases, to believe the valve was dysfunctional and resulting in -with hindsight- unnecessary re-operations. Based on these two clinical reports Medtronic communicated worldwide about the phenomenon of intermittent regurgitation. Since that communication there have been no further reports of unnecessary reoperations.”

Zelf controleren

Voor dit citaat hoeft niemand Bron op zijn woord te geloven. Het document dat rechtspraak.nl publiceerde, staat op https://bron.page/api/v1/blobs/65ade95e75f26496e08a7f4d1c0e507ec65d98e8b3f0d5e0af8d8abdab903b49. Wie het ophaalt en de SHA-256 ervan berekent, krijgt 65ade95e75f26496e08a7f4d1c0e507ec65d98e8b3f0d5e0af8d8abdab903b49, en de alinea staat in dat document op bytes 17864 tot 19031.